With four available output beam wavelenghts and a wide range of repetition rates (with single-shot, burst and continuous operating modes), the Tempest provides a complete solution for many applications. |
|
Ease of Use
Reliable Operation
Exceptional Performance
Application Flexibility
Options
|
Tempest is easy to install and easy to operate. The laser requires no external cooling water, and it operates on single-phase 110/220 VAC power. A power supply-mounted key switch provides basic electrical control of Tempest. Once in the "on" position, all system functions are activated. The control panel, which can be located as far as 10 ft. (3m) from the power supply, provides complete system control. Included on the control panel is a trigger control to select single-shot, fixed, variable or external TTL-triggered operation. A potentiometer controls the repetition rate, variable up to the maximum for the model. The laser features synchronization options, with I/O BNCs located on the rear of the power supply. The external fire sequence is initiated by sending a TTL pulse to trigger the flashlamp. The Q-switch can be fired internally or by a second TTL pulse, providing precise control of the laser pulse. Outputs for the lamp and Q-switch provide TTL pulses to trigger interfacing equipment. The configuration of Tempest is very flexible requiring simple changes to switch from fundamental to optional harmonic wavelengths. The optional attenuator provides variable energy without affecting beam characteristics. Ergonomic design makes Tempest easy to set up and easy to operate. |
|
|
Tempest 10 | Tempest 20 | Tempest 30 | Tempest 300 | |||
| Repetition Rate (Hz) | 10 | 20 | 30 | 10 | |||
| Energy1 (mJ) |
1064 nm | 200 | 200 | 180 | 300 | ||
| 532 nm | 100 | 100 | 90 | 180 | |||
| 355 nm | 50 | 50 | 40 | 75 | |||
| 266 nm | 30 | 30 | 20 | 40 | |||
| Energy Stability2 (±%) |
1064 nm | 2 | 2 | 2 | 2 | ||
| 532 nm | 3.5 | 3.5 | 4 | 4 | |||
| 355 nm | 8 | 8 | 9 | 9 | |||
| 266 nm | 9 | 9 | 10 | 10 | |||
| Pulse Width3 (ns) | 3 - 5 | 3 - 5 | 3 - 5 | 3 - 5 | |||
| Divergence4 (mrad) | < 1 | < 1 | < 1 | < 1 | |||
| Beam Pointing Stability (µrad) | < 100 | < 200 | < 250 | < 200 | |||
| IR Beam Quality5 (TDL) | 1.5 | 1.5 | 2 | 1.5 | |||
| Jitter (±ns) 6 | 0.5 | 0.5 | 0.5 | 0.5 | |||
| Beam Diameter @ 1064 nm | 5 mm | 5 mm | 5 mm | 6 mm | |||
|
1) Optical losses due to optional attenuator will reduce maximum energy by 10% 2) Pulse-to-pulse for 98% of shots after 30 minutes warm up 3) Full width half maximum 4) Full angle for 86% of the energy, at 1/e2 point 5) Times diffraction limited, at 1/e2 point 6) From Q-Switch synch out to light pulse, 98% of 1000 shots |
|||||||
|
|
Laser Head Tempest 10,20,&30 |
Laser Head Tempest 300 |
Laser Head Base Plate |
Power Supply | Remote Control Panel | |
| Length | 13.4" / 34 cm2 | 15.0" / 38.0 cm2 | 7.0" / 17.8 cm | 20.5" / 52.1 cm2 | 6.3" / 16.0 cm | |
| Width | 7.0" / 17.8 cm | 7.0" / 17.8 cm | 9.0" / 22.9 cm | 8.7" / 22.1 cm | 8.2" / 20.8 cm | |
| Height | 3.5" / 8.9 cm | 3.5" / 8.9 cm | 0.5" / 1.3 cm | 15.7" / 39.9 cm | 3.5" / 8.9 cm | |
| Weight | 12 lbs / 5.5 kg | 15 lbs / 6.8 kg | 5 lbs / 2.3 Kg | 55 lbs / 25 kg | 5 lbs / 2.3 kg | |
| Umbilical Length | 8 ft / 2.4 m | 8 ft / 2.4 m | 10 ft / 3 m | |||
|
1) Nominal values only, contact NWR representative for full specifications 2) Add 5.0" / 13.0 cm for umbilical bend radius |
||||||
Operating Requirements |
Temperature |
70° ± 10° F (21° ± 5° C) | |
| Relative Humidity |
20 – 80% non-condensing | |
| Voltage |
95-120 or 200-250 VAC, 50/60 Hz | |
| Power |
600 Watts |
| © 2005 New Wave Research | Terms of Use | Privacy Policy | Site Map |